Product categories
MFC2000 Gas Mass Flow Controller
Thermal MEMS mass flow controller
Full scale 50 mLn/min to 200 Ln/min
Largest span in the MFC series
Analog 0 ~ 5 Vdc, 4 ~ 20 mA or RS485
Control pressure 0.1 ~ 0.8 MPa
7/16-20UNF to VCR 1/2” connections
The MFC2000 is a thermal MEMS mass flow controller for non-corrosive gases. It measures and controls mass flow directly, without temperature or pressure compensation, and holds the setpoint with a normally closed control valve. This is the widest span of the series: full scale from 0 ~ 50 mLn/min up to 0 ~ 5000 mLn/min, or from 0 ~ 1 up to 0 ~ 200 Ln/min.
It mounts inline in the gas line and accepts an analog 0 ~ 5 Vdc or 4 ~ 20 mA setpoint, or a digital RS485 command. Control pressure range is 0.1 ~ 0.8 MPa below 20 Ln/min and 0.1 ~ 0.7 MPa from 50 to 200 Ln/min, with a maximum operating differential pressure of 0.4 MPa and a typical settling time of 100 msec.
Three values define the choice: the full scale, the connection (7/16-20UNF to 9/16-18 UNF, compression fitting 1/8” ~ 1/2”, or VCR 1/4” ~ 1/2”), and the input signal. Accuracy is ±1.0% r.d. and ±0.2% f.s.; repeatability is ±0.3% r.d. and ±0.06% f.s.
Technical data as published in the manufacturer manual (section 7).
| Full-scale range | 0 ~ 50 mLn/min … 0 ~ 5000 mLn/min; 0 ~ 1 … 0 ~ 200 Ln/min |
|---|---|
| Accuracy | ± 1.0% r.d. (20 ~ 100% of full scale); ±0.2% f.s. (<20% of full scale) |
| Repeatability | ± 0.3% r.d. (20 ~ 100% of full scale); ±0.06% f.s. (<20% of full scale) |
| Turn-down ratio | 100:1 |
| Max control range | 102 %FS |
| Control pressure range | 0.1~0.8 (<20 Ln/min); 0.1~0.7 (50…200 Ln/min) MPa |
| Maximum operating differential pressure | 0.4 MPa |
| Setpoint source (input signal) | Analog: 0~5 Vdc or 4~20 mA; Digital: RS485 |
| Settling time (typical) | 100 msec |
| Working temperature | 0 ~ 55 °C |
| Humidity | <95, no condensation %RH |
| Burst pressure | 1.5 MPa |
| External leakage rate | ≤1 × 10⁻⁹ (He) Pa·m³/s |
| Power supply | (12 ~ 24) Vdc, 1A |
| Output signal | Analog: 0~5 Vdc or 4~20 mA; Digital: RS485 |
| Max null shift (analog) | ±30 mVdc |
| Control valve | Normally Closed (NC) |
| Electrical connector | DB9 |
| Mechanical connection | 7/16-20UNF…9/16-18 UNF; compression fitting 1/8”~1/2”; or VCR 1/4”~1/2” |
| Protection | IP40 |
| Storage temperature | -20 ~ +70 °C |
| Reference conditions | 0 °C, 101.325 kPa, air |
| Fluid compatibility | Non-corrosive |
| CE | EN61000-2; -3; -4 |
| Compliance | RoHS, REACH |