MF4600 Gas Mass Flow Meter
MEMS thermal mass flow meter, DN1 to DN8
Maximum overflow 2 to 200 SLPM by model
Accuracy ±(2.0+0.5FS) %
Analog 0.5 to 4.5 Vdc or 4 to 20 mA
RS485 Modbus half-duplex output
IP40 body with manual keypad control
MF4600 is a MEMS thermal gas mass flow meter built around four pipe diameters: MF4601 at DN 1.0 mm, MF4602 at DN 2.0 mm, MF4603 at DN 3.0 mm and MF4608 at DN 8.0 mm. Maximum overflow runs from 2 SLPM on the smallest body to 200 SLPM on the largest, and maximum flow change from 0.3 to 30 SLPM/sec, so the model is chosen by the flow the line actually carries rather than by a single series rating.
Accuracy is ±(2.0+0.5FS) % at a 50:1 turn-down ratio, with a 1.0 MPa maximum pressure and -10 to +55 °C working temperature. Readings are available on analog 0.5 to 4.5 Vdc or 4 to 20 mA, and digitally on RS485 Modbus half-duplex through an AMPMODU MTE 5-position connector. The front face carries an LCD with two indicators and a three-key keypad; control is manual. Power is 8 to 24 Vdc at 50 mA and the body is rated IP40.
Selection rests on three values: the DN that matches the fitting, the full-scale flow rate for that DN, and whether the controller takes the analog or the Modbus output. Mechanical connection and full-scale range are both tabulated in section 6 of the manual.
Technical data as published in the manufacturer manual (section 7).
| Full-scale flow range | See Table 6.1 |
|---|---|
| Accuracy | ±(2.0+0.5FS) % |
| Repeatability | (0.5+0.15FS) % |
| Turn-down ratio | 50:1 |
| Working temperature | -10 ~ +55 °C |
| Maximum pressure | 1.0 MPa |
| Humidity | <95, no condensation %RH |
| Power supply | 8 ~ 24 (50 mA) Vdc |
| Digital output | RS485 Modbus half-duplex |
| Analog output | 0.5 ~ 4.5 Vdc / 4 ~ 20 mA |
| Electrical connector | AMPMODU MTE 5 positions |
| MENU access | 3 key – front face keyboard/digital |
| Display | Instant flow rate, totalizer, or accumulated flow rate with LED & 2 indicators |
| Control | Manual |
| Mechanical connection | See Table 6.1 |
| Protection | IP40 |
| Storage temperature | -20 ~ +70 °C |
| Reference conditions | 20°C, 101.325 kPa, air |
| Fluid compatibility | Non-corrosive |
| CE | EN61326-1; -2; -3 |
| Maximum overflow (MF4601, DN 1.0 mm) | 2 SLPM |
| Maximum overflow (MF4602, DN 2.0 mm) | 6 SLPM |
| Maximum overflow (MF4603, DN 3.0 mm) | 30 SLPM |
| Maximum overflow (MF4608, DN 8.0 mm) | 200 SLPM |
| Maximum flow change (MF4601, DN 1.0 mm) | 0.3 SLPM/sec |
| Maximum flow change (MF4602, DN 2.0 mm) | 1 SLPM/sec |
| Maximum flow change (MF4603, DN 3.0 mm) | 4 SLPM/sec |